Wika DPT-10 Manuale Utente Pagina 18

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 96
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 17
+
1
2 2
3
5 5
6
4 4
Fig. 12: Measurement setup, ow measurement in liquids
1 Orice or impact pressure probe
2 Blocking valves
3 DPT10
4 Precipitator
5 Drain valves
6 3-fold valve block
4.4 Measurement setup level
à Mount DPT10 below the lower measurement connection so that
the eective pressure lines are always lled with liquid
à Minus side is open to the atmospheric pressure
In open vessels with
eective pressure line
18 Dierential pressure transmitter DPT10 - metal measuring diaphragm Probus PA
4 Mounting
37245-EN-100120
Vedere la pagina 17
1 2 ... 13 14 15 16 17 18 19 20 21 22 23 ... 95 96

Commenti su questo manuale

Nessun commento