Density measurement
=
p
h
g
+
h
Fig. 5: Density measurement with DPT10, h = dened mounting distance, Δp =
dierential pressure, ρ = density of the medium, g = acceleration of gravity
1 DPT10
Interface measurement
1
+
h
2
3
Fig. 6: Interface measurement with DPT10
1 DPT10
2 Liquid with highest density
3 Liquid with lowest density
A metallic measuring cell is used as sensor element. The process
pressures are transmitted via separating diaphragms and lling oils to
the resistance measuring bridge (semi-conductor technology). The
dierential-pressure dependent change of the bridge voltage is
measured, further processed and converted into a corresponding
output signal.
Functional principle
10 Dierential pressure transmitter DPT10 - metal measuring diaphragm Probus PA
3 Product description
37245-EN-100120