11498171.01 08/2009 GB
13. Storage of the system
Before the system is stored it must be clean and free of dirt and debris.
This is particularly important if the medium is a health hazard such as a
corrosive, toxic, carcinogenic, radioactive, etc.
The storage place must satisfy the following conditions:
Ambient temperature: 0 to 70 °C
Humidity: 35 to 85 % relative humidity without condensation
Avoid the following influences:
Direct sunlight or vicinity to hot objects
Mechanical vibration
Soot, steam, dust and corrosive gasses
Explosion-hazard environment, inflammable atmosphere
The system should be stored in its original transport box, in a place that meets the conditions
listed above.
Follow the following instructions to avoid damage.
1. Wrap the system in anti-static plastic foil.
2. Using the insulating material, place the system in the box.
3. If the system is stored for a longer time (more than 30 days) add a bag with desiccant to the
box.
Warning