Pneumatic High-Speed Pressure Controller CPC3000
WIKA Operating Instruction Pneumatic High-Speed Pressure Controller ∙ Version 1.146
11498171.01 08/2009 GB
Command Data Response/Function
Rdecpt? <sp>n<cr><lf> Returns the number of rate decimal
points for the active channel. (see
Resolution)
Release? 15 char string.
Ex:
Release? Test_stand_1
Returns:
<sp>(Yes or No), CCC…
CCCcr><lf>
This command is used to release
control of the instrument in a multiple
computer environment.
Yes if release is successful.
No if instrument is being controlled
with another computer.
CCC… = name of controlling computer
or AVAILABLE
See: Acquire? and Unlock
Resolution n Sets the number of significant digits.
See decpt.
Resolution? <sp>n<cr><lf> Returns the number of significant
digits. See decpt.
Rfilter Value in % Sets the % of the rate filter
Rfilter? <sp>n.nnnnne+nn<cr><lf> Returns the rate filter.
Rsetpt Value in current units Sets the rate setpoint
Rsetpt? <sp>n.nnnnne+nn<cr><lf> Returns the rate setpoint
Rfreq Value in frequency Sets rate Butterworth corner frequency.
Rfreq? <sp>n.nnnnne+nn<cr><lf> Returns rate Butterworth corner
frequency.
Rwindow Value in current units Sets rate exponential filter window.
Rwindow? <sp>n.nnnnne+nn<cr><lf> Returns rate exponential filter window.
Sbaud 9600, 19200, 38400, 57600 Sets the serial baud rate.
Sbaud? <sp>XXXX<cr><lf> Returns the serial baud data.
Sdata 7 or 8 Sets the serial data bits.
Sdata? <sp>X<cr><lf> Returns the serial data bits number.
Sensor C, X Sets the active sensor where C =
Primary or Secondary and X is the
turndown.
Sensor? <sp>C,X<cr><lf> Returns active sensor as above.
Sensorid? <sp>Mensor QRS,SN XXXXXX,VER
V.VV
Returns the active sensor’s serial
number and firmware version.
Setpt value inside upper and lower limits
and inside the range of the active
sensor and turndown.
Sets the control setpoint for the instru-
ment.
Setpt? <sp>XXXXXXX<cr><lf> Returns the control setpoint in current
units.
Setpt% Value in % of current range Sets the control setpoint in % of
current range.
Setptpct Value in % of current range Sets the control setpoint in % of
current range.
Setptpct? <sp>n.nnnnne+nn<cr><lf> Returns the current setpoint in % of
current range.