Wika CPC2000 Manuale Utente Pagina 24

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 132
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 23
24 WIKA operating instructions, model CPC2000
11155094.02 07/2012 GB/D/F/E
GB
6. Commissioning, operation
6.5.1.2 Unit
Here the unit, in which the low-pressure controller will display the measured value on the
screen, can be selected. Dependent upon the measuring range of the instrument, various
units are not available since they could not be displayed.
6.5.1.3 Steps
Here the step size for the STEP function, in x % of the set range, can be changed. Possi-
ble settings are from 1 … 50 %.
6.5.1.4 Mode
Here the operating mode of the controller can be set. The following modes can be
selected:
Measure (MEAS): The controller measures the pressure in the set pressure unit
Control (CTRL): A set point is automatically regulated.
Automatic (AUTO): A stored test routine will be carried out.
6.5.1.5 Language
Here the language in which the menu will be displayed can be set. There is the possibility
to choose between German, English, Spanish and French.
6.5.1.6 Settings
ZERO: Switching on and o of the automatic zero point adjustment following exit from
the menu. It is even possible to set at what intervals the instrument will automatically
perform a zero point adjustment.
RS232/USB: Selection of the appropriate serial interface
DISPLAY: Brightness setting
AUTO-MODE: Setting of the pressure range, the pressure unit and number of steps
the pressure range will be divided up into. The STEPS UP/STEPS DOWN can be
assigned dierent values.
INFO: Basic information on the instrument
Vedere la pagina 23
1 2 ... 19 20 21 22 23 24 25 26 27 28 29 ... 131 132

Commenti su questo manuale