11WIKA operating instructions, model CEP3000
14036068.02 08/2012 GB/D
GB
3. Specications
Input and output
signals
Measuring
range
Absolute measurement
uncertainty (all errors
incl.)
Resistance thermometer
1)
Pt100 (385) -200.0 ... +800.0 °C 0.33 °C
Pt100 (3926) -200.0 ... +630.0 °C 0.3 °C
Pt100 (3916) -200.0 ... +630.0 °C 0.3 °C
Pt200 (385) -200.0 ... +630.0 °C 0.8 °C
Pt500 (385) -200.0 ... +500.0 °C
500.0 ... 630.0 °C
0.3 °C
0.4 °C
Pt1000 (385) -200.0 ... +100.0 °C
100.0 ... 680.0 °C
0.2 °C
0.3 °C
Ni120 (672) -80.0 ... +260.0 °C 0.2 °C
Cu 10 -100.0 ... +250.0 °C 2.2 °C
Cu 50 -180.0 ... +200.0 °C 0.5 °C
Cu 100 -180.0 ... +200.0 °C 0.3 °C
YSI 400 15.0 ... 50.0 °C 0.1 °C
Resistance Excitation current
Output 5.0 ... 400.0 Ω
5.0 ... 400.0 Ω
401 ... 1,500 Ω
1,501 ... 3,200 Ω
0.15 Ω
0.1 Ω
0.5 Ω
1.0 Ω
0.1 ... 0.5 mA
0.5 ... 3 mA
0.05 ... 0.8 mA
0.05 ... 0.4 mA
Input 0.00 ... 400.00 Ω
401.0 ... 1,500.0 Ω
1,501.0 ... 3,200.0 Ω
0.1 Ω
0.5 Ω
1.0 Ω
1) Absolute measurement uncertainty based on a 4-wire circuit
For further specications see WIKA data sheet CT 82.01 and the order documentation.