9WIKA safety manual temperature transmitter T32.xS
11583631.03 08/2014 GB/D/F/E
Rail mounting version, model T32.3S
2.3 Restrictions to operating modes
WARNING!
Under the following operating conditions, the safety function of the
instrument is not guaranteed:
During conguration
When the write protection is deactivated
When the HART
®
multi-drop mode is activated
Measured value transmission via HART
®
protocol
During a simulation
During the proof test
Pin assignment
Model
with SIL: T32.1S.0IS-S
without SIL: T32.1S.0IS-Z
Date of manufacture
(year-month)
SIL version
(only for SIL)
2. Safety
Vedere la pagina 8