Wika BCS10 Manuale Utente Pagina 10

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 108
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 9
10
WIKA operating instructions calibration system, model BCS10
14051046.01 10/2012 GB/D/F/E
GB
3. Specications
Digital indicator
Type of indication: 7-segment LCD
Digits: 5 ½-digit
Resolution: 0.001
Character size: 16.53 mm (0.65")
Bargraph display: 20 segment bargraph, 0 ... 100 %
Possible pressure units: bar, psi, kg/cm
2
, kPa, MPa and 15 other units
Operating conditions
Ingress protection: IP 65
Ambient temperature: -10 ... +55 °C
Medium temperature: 0 ... +55 °C
Storage temperature: -20 ... +70 °C
Relative humidity: < 95 % r.H. (non-condensing)
Communication
Interface: USB, RS-232
Functions
Overpressure protection: Adjustable overpressure valve
Pressure ne-adjustment: By means of the ne-adjustment valve
Measuring rate: 30/min. to 10/sec. selectable
Memory: MIN/MAX values
Integrated data logger
Data logger: Cyclic data logger: automatic recording of up to 8.500
values
Cycle time: selectable from 1 ... 3.600 sec. in the following
steps
1 s, 2 s, 5 s, 10 s, 30 s, 1 min, 2 min, 5 min, 10 min,
30 min and 1 h
CPGLog data-logger evaluation software is needed to
use the data logger function.
The order number of the CPGlog software is given in the
data sheet SP 60.08.