Wika TR31 Manuale Utente Pagina 22

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 56
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 21
GB
14096788.01 07/2014 GB/D
22 WIKA operating instructions models TR31-3, TR31-K
6. Commissioning, operation
Hysteresis on return to the measuring span
After the linear error limits have been exceeded, on return to the
measuring span, a hysteresis of 0.1 mA must be passed. This
hysteresis prevents the transmitter from jumping back and forth
between error and measuring mode.
Fault signal
Return from fault
signal
Return from fault
signal
3.8 ...
4 mA
20 ... 20.5 mA
4 ... 20 mA
3.8 mA 4 mA 20 mA 20.5 mA
3.9 mA 20.4 mA
Fault signal
Error limits and permissible measuring range
Vedere la pagina 21
1 2 ... 17 18 19 20 21 22 23 24 25 26 27 ... 55 56