31WIKA operating instructions, model CPH7600
14051628.01 11/2012 GB/D
GB
6. Commissioning, operation
6.9 Measuring voltage
For voltage measurement, use the connections on the front of the calibrator. Select the
VOLTS function on one of the displays. The calibrator can measure up to DC 30 V.
The display will indicate "OL" when the measured voltage exceeds the
nominal range of voltage (30 V).
Test item
4 ... 20 mA
DC 24 V
Test item
Up to DC 30 V
Test item
Voltage supply
4 ... 20 mA
Vedere la pagina 30