52 WIKA operating instructions ProcessCalibrator, model CPH6000
11069023.04 11/2012 GB/D
GB
Customer-specic unit; with respect to bar
(input via numeric keypad)
Display resolution in operating mode
(back with )
Select and conrm (standard units) via
Short info:
XXX
Current cursor position; alter via
◁▷ ◀▶
Parameter selection from list or menu via
0.00 Parameter input via numeric keypad
ENTER
Input conrmation
CLEAR
Clear entry
In order to switch the instrument into "SWITCH-TEST" mode, follow the instructions on
the previous page.
The switch-test function is not suitable for electronic switches (e.g.
PNP- or NPN switches), but is only for mechanical, potential-free
switches.
The following is a more detailed explanation of point "2. Preparation for SWITCH-
TEST"
a) Voltage supply for test item (on/o) [If no external supply is required for the
test item, "OFF" should be selected to conserve energy]
b) Unit and resolution (submenu)
7. Commissioning, operation
The calculated measured values of the two switching points and the
hysteresis can be reset by pressing the "0" key.